Product Details:
|
Flow Rate Range: | 0.1~600ml/h Micro Mode: 0.1-99.9 Ml/hr In 0.1 Ml Increments | Volume Infused: | 0~36000ml |
---|---|---|---|
Applicable Infusion Set: | Support A Wide Variety Brands Of Infusion Set Including Major Manufacturers & Selected Local Brands | Bolus Rate: | 300 ~ 600 Ml/h |
Alarms: | Visual And Audible Alarms:Door Open,Air-In-Line, Occlusion,Infusion Completion, Infusion Near Over, Start Reminder Function, Low Battery,Battery Depleted,malfunction Etc. | History Records: | 2000 History Records |
Dimensions: | 145×100×120 Mm(length×width×height) | Weight: | ≤1.4kg |
High Light: | 36000ml Veterinary Infusion Pump,600ml/h Veterinary Infusion Pump |
Veterinary Infusion Pump Medical Infusion Pump Animal Vet Use With Volume Infused of 0~36000ml
Feature & Functions
1 | Compact design |
The compact and light weight design saves space Large & colorful LCD display |
2 | Free flow protection |
The built-in flow stop feature helps reduce the risk of accidental free flow prevent when pump door is open. |
3 | Anti-bolus function |
Diminishes the volume of unwanted Bolus injected to the patient after removal of the occlusion cause. Anti-reverse rotation detection |
4 | Panel lock | The panel lock feature helps prevent unauthorized changes of any instrument setting |
5 | Versatile functions |
2000 history records Real time display RS232 interface (optional) 90°rotation pole clamp convenient for horizontal bar, vertical bar, vertical IV pole & platform use |
6 | User-friendly operation |
Soft key design, easy to operate Load directly the last infusion rate and volume limit Adjustable buzzer volume Top indicator light reflects the operation status as well as alarm case Night mode to ensure good rest of patients removable pump body for easy cleaning |
Specification
1 |
Flow rate range |
0.1~1200ml/h Micro mode: 0.1-99.9 ml/h in 0.1 ml increments |
2 |
Volume to be infused (VTBI) |
0~9999ml |
3 |
Volume infused (∑) |
0~36000ml |
4 |
Infusion accuracy |
±5% |
5 |
Applicable infusion set |
Support a wide variety brand of infusion set including major manufactures & selected local brands |
6 |
KVO rate |
1~5ml/h |
7 |
Bolus rate |
300~600ml/h |
8 |
Purge |
600ml/h |
9 |
Air bubble detection |
Ultrasound sensor detection with four levels of air in line alarm adjustable |
10 |
Occlusion pressure |
3 adjustable occlusion pressure settings: low, middle and high |
11 |
Alarms |
Visual and audible alarms: Door open, Air-In-Line, Occlusion, Infusion completion, infusion near over, start reminder function, low battery, battery depleted, malfunction etc |
12 |
Computer interface |
Rs232 (optional) |
13 |
History records |
2000 history records |
14 |
Power supply |
AC:100~240V,50/60Hz DC:12V±1V |
15 |
Battery |
Rechargeable lithium polymer battery, 7.4V, 1900mAh Can operate for more than 3 hours after fully charged at 25ml/h |
16 |
Battery recharge time |
10h with power on, 3h with power off |
17 |
Power consumption |
25VA |
18 |
Classification |
class I, type CF |
19 |
Waterproof level |
IP * 3 |
20 |
Mode of operation |
continuous |
21 |
Dimensions |
145 * 100 *120 mm (length * width * height) |
22 |
Weight |
≤1.4kg |
23 |
Operating conditions |
Temperature:5°~40°,relative humidity:10~95%, Atmosphere pressure: 86~106kPa |
24 |
Transport conditions |
Temperature:-20°~ +60°, relative humidity: 10~95%, Atmosphere pressure: 50~106kPa |
25 |
Storage conditions |
Temperature:-20°~ +45°,relative humidity:10~95%, Atmosphere pressure: 50~106kPa |
Contact Person: Mrs. Helen Bo